O&M Manual
Rev-H (8/11)
Part 7 - Maintenance
The A15/76 Turbidity Monitor will generally provide unattended operation over long periods of
time. With proper care, the system should continue to provide measurements indefinitely. For reliable
operation, maintenance on the system must be done on a regular schedule. Keep in mind that preventive
maintenance on a regular schedule is much less troublesome than emergency maintenance that always
seems to come at the wrong time.
Sensor Maintenance
Virtually all of the maintenance required for operation of the turbidity monitor is sensor related.
The electronics are generally trouble free. They are burned in at the factory and will likely have a problem
only if random component failure occurs.
Sensor maintenance is required for accurate measurements. The primary requirement is simply
to keep the sensor clean. Wiping the sensor with a soft cloth is normally sufficient. Calibration should be
checked at monthly intervals after initial installation.