Model A15/76 Turbidity Monitor Home Office European Office Analytical Technology, Inc. ATI (UK) Limi
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 10 O&M Manual Rev-H (8/11) A special handrail mounting kit (part
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 11 O&M Manual Rev-H (8/11) The majority of turbidity applications
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 12 O&M Manual Rev-H (8/11) Figure 7 below shows the recommended i
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 13 O&M Manual Rev-H (8/11) Turbidity sensors may also be installe
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 14 O&M Manual Rev-H (8/11) Figure 9 - Flow "T" Assemb
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 15 O&M Manual Rev-H (8/11) For applications where a submersible
16 O&M Manual Rev-H (8/11) Part 3 – Electrical Connections Wiring connections to the A15/76 are made to the back of the monitor. The termina
ATI A15/76 Turbidity Monitor Part 3 – Electrical Connections 17 O&M Manual Rev-H (8/11) DC Powered Monitor Model A15/76 Monitor
ATI A15/76 Turbidity Monitor Part 3 – Electrical Connections 18 O&M Manual Rev-H (8/11) Figure 13 - Remote Sensor Junction Box
19 O&M Manual Rev-H (8/11) Part 4 – Start-Up Prior to operation, recheck electrical connections to be sure everything is in accordance with F
2 O&M Manual Rev-H (8/11) PRODUCT WARRANTY Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the
ATI A15/76 Turbidity Monitor Part 4 – Startup 20 O&M Manual Rev-H (8/11) Front Panel Lock The key swi
21 O&M Manual Rev-H (8/11) Part 5 - Configuration The electronics for the Turbidity system are designed to be as flexible as possible. A num
ATI A15/76 Turbidity Monitor Part 5 – Configuration 22 O&M Manual Rev-H (8/11) 8. Small S RT: 120s This routine
ATI A15/76 Turbidity Monitor Part 5 – Configuration 23 O&M Manual Rev-H (8/11) 19. AL TYPE: CONT. This routine
24 O&M Manual Rev-H (8/11) Part 6 - Operation Turbidity monitors will start to measure aqueous samples as soon as power is applied. Calibr
ATI A15/76 Turbidity Monitor Part 6 – Operation 25 O&M Manual Rev-H (8/11) Manual Low Level Offset Adjustmen
ATI A15/76 Turbidity Monitor Part 6 – Operation 26 O&M Manual Rev-H (8/11) 1. Refer to Figure 7 on page 11.
ATI A15/76 Turbidity Monitor Part 6 – Operation 27 O&M Manual Rev-H (8/11) Controls & Alarm Outputs T
ATI A15/76 Turbidity Monitor Part 6 – Operation 28 O&M Manual Rev-H (8/11) The alarm relay will always acti
ATI A15/76 Turbidity Monitor Part 6 – Operation 29 O&M Manual Rev-H (8/11) Auto/Manual/ Simulate Operating M
3 O&M Manual Rev-H (8/11) TABLE OF CONTENTS Unpacking ...
ATI A15/76 Turbidity Monitor Part 6 – Operation 30 O&M Manual Rev-H (8/11) Check Signal Display The check
31 O&M Manual Rev-H (8/11) Part 7 - Maintenance The A15/76 Turbidity Monitor will generally provide unattended operation over long periods of
32 O&M Manual Rev-H (8/11) SPARE PARTS LIST PART NO. DESCRIPTION 00-0804 Complete turbidity monitor electronics, panel mount 03-0181 NE
4 O&M Manual Rev-H (8/11) TABLE OF FIGURES Figure 1 - Turbidity System w/Submersion Sensor ...
5 O&M Manual Rev-H (8/11) Unpacking When you receive your Turbidity Monitor, open the shipping package and inspect the contents to be sure
6 O&M Manual Rev-H (8/11) Part 1 - Introduction The Model A15/76 is an on-line monitoring system designed for the continuous measurement of t
ATI A15/76 Turbidity Monitor Part 1 - Introduction 7 O&M Manual Rev-H (8/11) Specifications Electronics Unit Range: 4/400 NTU
8 O&M Manual Rev-H (8/11) Part 2 – Mechanical Installation The main installation requirements for ATI’s Turbidity Monitor are mechanical insta
ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation 9 O&M Manual Rev-H (8/11) The optional NEMA 4X wall mount enclosure