ATI A15/76 Turbidity Monitor Part 2 – Mechanical Installation
11
O&M Manual
Rev-H (8/11)
The majority of turbidity applications require the use of a flowcell assembly. This method is best
when monitoring very low turbidity values, such as filter effluent. The flowcell used with ATI’s turbidity
system is designed to eliminate ambient light effects, and to allow sample pressure to be maintained
through the flowcell minimizing air bubble formation that can cause erroneous turbidity measurements. A
bracket is supplied for mounting the flowcell to any flat surface. Figure 5 and 6 show details of the flowcell
mounting. The sample outlet must be above the sample inlet for proper operation.
Figure 5 - Flowcell Dimensions
Figure 6 - Exploded Flowcell & Sensor Assembly