ATI A15/66 Residual Sulfite Monitor
9
O&M Manual
Rev-W (8/10)
ELECTRICAL CONNECTIONS
AC power must be connected to both the monitor and the chemistry module. Refer to Figure 5 below for
the proper connections. A power cord is supplied with each unit for AC connection and a 6- conductor
interconnect cable (20 ft.) is supplied for connection of the chemistry module to the monitor. Note that the
terminal blocks at the rear of the monitor unplug for wiring convenience. Figure 6 provides the
interconnection wiring diagram between the chemistry module and the monitor. The chemistry module
can be located up to 100 feet from the monitor provided that connection is made with a single length of
#31-0037 cable. Cable must be ordered separately.
TB1 TB2
1 - AC POWER - NEUTRAL 14 - MA OUTPUT - (+)
2 - AC POWER - 110V - HOT 15 - MA OUTPUT - (+) (OPTIONAL DUAL)
3 - AC POWER - 220V - HOT 16 - MA OUTPUT - (-)
4 - AC - GROUND 17 - SENSOR - WHITE
5 - RELAY A - NORMALLY OPEN 18 - SENSOR - BROWN OR ORANGE
6 - RELAY A - COMMON 19 – SENSOR - BLUE
7 - RELAY A - NORMALLY CLOSED 20 - UNUSED
8 - RELAY B - NORMALLY OPEN 21 - UNUSED
9 - RELAY B - COMMON 22 - UNUSED
10 - RELAY B - NORMALLY CLOSED 23 – SENSOR – RED
11 - ALARM RELAY C/D - NORMALLY OPEN 24 – SENSOR – BLACK
12 - ALARM RELAY C/D - COMMON 25 – SENSOR – GREEN
13 - ALARM RELAY C/D - NORMALLY OPEN
Figure 5 - Electrical Connections